Research Article | Open Access

Running-in friction of hip joint replacements can be significantly reduced: The effect of surface-textured acetabular cup

David NEČAS1( )Hatsuhiko USAMI2Tatsuya NIIMI3Yoshinori SAWAE4Ivan KŘUPKA1Martin HARTL1
Faculty of Mechanical Engineering, Brno University of Technology, Brno 61669, Czech Republic
Faculty of Science and Technology, Meijo University, Nagoya 468-8502, Japan
Kanefusa Corporation, Nagoya 480-0192, Japan
Faculty of Engineering, Kyushu University, Fukuoka 819-0395, Japan
Show Author Information

Abstract

Hip joint replacements represent the most effective way of treatment for patients suffering from joint diseases. Despite the rapid improvement of implant materials over the last few decades, limited longevity associated with wear-related complications persists as the main drawback. Therefore, improved tribological performance is required in order to extend the service life of replacements. The effect of surface texturing of ultra- high molecular weight polyethylene (UHMWPE) acetabular cup was investigated in the present study. Unique tilling method was utilized for manufacturing the dimples with controlled diameter and depths on the contact surface of the cup. The experiments with four commercial femoral components and two model lubricants were realized. The main attention was paid to a coefficient of friction considering the differences between the original and the dimpled cups. Results showed remarkable lowering of friction, in general. Focusing on the simulated human synovial fluid, friction was reduced by 40% (alumina ceramic), 38.8% (zirconia toughened ceramic), 25.5% (metal), and 9.9% (oxinium). In addition, the dimples helped to keep the friction stable without fluctuations. To conclude, the paper brings a new insight into frictional behaviour of the hip replacements during running-in phase which is essential for overall implant lifespan. It is believed that proper surface texturing may rapidly improve the life quality of millions of patients and may lead to considerable financial savings.

References

[1]
J Gallo, S B Goodman, Y T Konttinen, M A Wimmer, M Holinka. Osteolysis around total knee arthroplasty: A review of pathogenetic mechanisms. Acta Biomater 9(9): 8046-8058 (2013)
[2]
J Gallo, J Vaculova, S B Goodman, Y T Konttinen, J P Thyssen. Contributions of human tissue analysis to understanding the mechanisms of loosening and osteolysis in total hip replacement. Acta Biomater 10(6): 2354-2366 (2014)
[3]
P N Tandon, S Jaggi. Wear and lubrication in an artificial knee joint replacement. Int J Mech Sci 23(7): 413-422 (1981)
[4]
C B Rieker. Tribology of total hip arthroplasty prostheses: What an orthopaedic surgeon should know. EFORT Open Rev 1(2): 52-57 (2016)
[5]
J R Penmetsa, P J Laz, A J Petrella, P J Rullkoetter. Influence of polyethylene creep behavior on wear in total hip arthroplasty. J Orthop Res 24(3): 422-427 (2006)
[6]
J Zeman, M Ranuša, M Vrbka, J Gallo, I Křupka, M Hartl. UHMWPE acetabular cup creep deformation during the run-in phase of THA’s life cycle. J Mech Behav Biomed Mater 87: 30-39 (2018)
[7]
C Myant, P Cann. On the matter of synovial fluid lubrication: Implications for Metal-on-Metal hip tribology. J Mech Behav Biomed Mater 34: 338-348 (2014)
[8]
M Vrbka, D Nečas, M Hartl, I Křupka, F Urban, J Gallo. Visualization of lubricating films between artificial head and cup with respect to real geometry. Biotribology 1-2: 61-65 (2015)
[9]
C W Myant, P Cann. The effect of transient conditions on synovial fluid protein aggregation lubrication. J Mech Behav Biomed Mater 34: 349-357 (2014)
[10]
D Nečas, M Vrbka, F Urban, J Gallo, I Křupka, M Hartl. In situ observation of lubricant film formation in THR considering real conformity: The effect of diameter, clearance and material. J Mech Behav Biomed Mater 69: 66-74 (2017)
[11]
D Nečas, M Vrbka, A Galandáková, I Křupka, M Hartl. On the observation of lubrication mechanisms within hip joint replacements. Part I: Hard-on-soft bearing pairs. J Mech Behav Biomed Mater 89: 237-248 (2019)
[12]
D Nečas, M Vrbka, D Rebenda, J Gallo, A Galandáková, L Wolfová, I Křupka, M Hartl. In situ observation of lubricant film formation in THR considering real conformity: The effect of model synovial fluid composition. Tribol Int 117: 206-216 (2018)
[13]
F Ali, I Křupka, M Hartl. Reducing the friction of lubricated nonconformal point contacts by transverse shallow micro- grooves. Proc Inst Mech Eng Part J J Eng Tribol 229(4): 420-428 (2015)
[14]
C Gachot, A Rosenkranz, S M Hsu, H L Costa. A critical assessment of surface texturing for friction and wear improvement. Wear 372-373: 21-41 (2017)
[15]
H Usami, T Sato, Y Kanda, S Nishio. Applicability of Interrupted Micro Cutting Process “Tilling” as Surface Texturing. Key Eng Mater 749: 241-245 (2017)
[16]
H Ito, K Kaneda, T Yuhta, I Nishimura, K Yasuda, T Matsuno. Reduction of polyethylene wear by concave dimples on the frictional surface in artificial hip joints. J Arthroplasty 15(3): 332-338 (2000)
[17]
H Sawano, S Warisawa, S Ishihara. Study on long life of artificial joints by investigating optimal sliding surface geometry for improvement in wear resistance. Precis Eng 33(4): 492-498 (2009)
[18]
A Borjali, J Langhorn, K Monson, B Raeymaekers. Using a patterned microtexture to reduce polyethylene wear in metal-on- polyethylene prosthetic bearing couples. Wear 392-393: 77-83 (2017)
[19]
J Langhorn, A Borjali, E Hippensteel, W Nelson, B Raeymaekers. Microtextured CoCrMo alloy for use in metal-on- polyethylene prosthetic joint bearings: Multi-directional wear and corrosion measurements. Tribol Int 124: 178-183 (2018)
[20]
M Cho, H J Choi. Optimization of surface texturing for contact between steel and ultrahigh molecular weight polyethylene under boundary lubrication. Tribol Lett 56(3): 409-422 (2014)
[21]
T Roy, D Choudhury, S Ghosh, A B Mamat, B Pingguan- Murphy. Improved friction and wear performance of micro dimpled ceramic-on-ceramic interface for hip joint arthroplasty. Ceram Int 41(1): 681-690 (2015)
[22]
D Choudhury, M Vrbka, A B Mamat, I Stavness, C K Roy, R Mootanah, I Krupka. The impact of surface and geometry on coefficient of friction of artificial hip joints. J Mech Behav Biomed Mater 72: 192-199 (2017)
[23]
M Vrbka, D Nečas, J Bartošík, M Hartl, I Křupka, A Galandáková, J allo. Determination of a Friction Coefficient for THA Bearing Couples. Acta Chir Orthop Traumatol Čech 82(5): 341-347 (2015)
[24]
L M Gao, P R Yang, I Dymond, J Fisher, Z M Jin. Effect of surface texturing on the elastohydrodynamic lubrication analysis of metal-on-metal hip implants. Tribol Int 43(10): 1851-1860 (2010)
[25]
L M Gao, Z K Hua, R Hewson, M S Andersen, Z M Jin. Elastohydrodynamic lubrication and wear modelling of the knee joint replacements with surface topography. Biosurf Biotribol 4(1): 18-23 (2018)
[26]
A Chyr, M F Qiu, J W Speltz, R L Jacobsen, A P Sanders, B Raeymaekers. A patterned microtexture to reduce friction and increase longevity of prosthetic hip joints. Wear 315(1-2): 51-57 (2014)
[27]
D Choudhury, D Rebenda, S Sasaki, P Hekrle, M Vrbka, M Zou. Enhanced lubricant film formation through micro- dimpled hard-on-hard artificial hip joint: An in-situ observation of dimple shape effects. J Mech Behav Biomed Mater 81: 120-129 (2018)
[28]
T E Stanton. Boundary lubrication in engineering practice. Engineer 29: 678-680 (1923)
[29]
M Ranuša, J Gallo, M Vrbka, M Hobza, D Paloušek, I Křupka, M Hartl. Wear analysis of extracted polyethylene acetabular cups using a 3D optical scanner. Tribol Trans 60(3): 437-447 (2017)
[30]
D Nečas, Y Sawae, T Fujisawa, K Nakashima, T Morita, T Yamaguchi, M Vrbka, I Křupka, M Hartl. The Influence of proteins and speed on friction and adsorption of metal/ UHMWPE contact pair. Biotribology 11: 51-59 (2017)
[31]
Y Sawae, T Murakami, J Chen. Effect of synovia constituents on friction and wear of ultra-high molecular weight polyethylene sliding against prosthetic joint materials. Wear 216(2): 213-219 (1998)
[32]
A Mavraki, P M Cann. Lubricating film thickness measurements with bovine serum. Tribol Int 44(5): 550-556 (2011)
[33]
J T Evans, R W Walker, J P Evans, A W Blom, A Sayers, M R Whitehouse. How long does a knee replacement last? A systematic review and meta-analysis of case series and national registry reports with more than 15 years of follow- up. Lancet 393(10172): 655-663 (2019)
[34]
F di Puccio, L Mattei. Biotribology of artificial hip joints. World J Orthop 6(1): 77-94 (2015)
[35]
G H Isaac. Life cycle aspects of total replacement hip joints. Tribol Interface Eng Ser 48: 147-160 (2005)
[36]
A J Rao, E Gibon, T Ma, Z Y Yao, R L Smith, S B Goodman. Revision joint replacement, wear particles, and macrophage polarization. Acta Biomater 8(7): 2815-2823 (2012)
[37]
R Lappalainen, S S Santavirta. Potential of coatings in total hip replacement. Clin Orthop Relat Res 430: 72-79 (2005)
[38]
D Choudhury, J M Lackner, L Major, T Morita, Y Sawae, A B Mamat, I Stavness, C K Roy, I Krupka. Improved wear resistance of functional diamond like carbon coated Ti-6Al- 4V alloys in an edge loading conditions. J Mech Behav Biomed Mater 59: 586-595 (2016)
[39]
H Haider, J N Weisenburger, F Namavar, K L Garvin. Why coating technologies for hip replacement systems, and the importance of testing them in vitro. Oper Tech Orthop 27(3): 152-160 (2017)
[40]
V K Balla, S Banerjee, S Bose, A Bandyopadhyay. Direct laser processing of a tantalum coating on titanium for bone replacement structures. Acta Biomater 6(6): 2329-2334 (2010)
[41]
S B Goodman, E Gómez Barrena, M Takagi, Y T Konttinen. Biocompatibility of total joint replacements: A review. J Biomed Mater Res A 90A(2): 603-618 (2009)
[42]
M N Rahaman, T Huang, B S Bal, Y Li. In vitro testing of Al2O3-Nb composite for femoral head applications in total hip arthroplasty. Acta Biomater 6(2): 708-714 (2010)
[43]
Ł Łapaj, J Markuszewski, J Wendland, A Mróz, M Wierusz- Kozłowska. Massive failure of TiNbN coating in surface engineered metal-on-metal hip arthroplasty: Retrieval analysis. J Biomed Mater Res B Appl Biomater 104(5): 1043-1049 (2016)
[44]
S Kurtz, K Ong, E Lau, F Mowat, M Halpern. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89(4): 780-785 (2007)
[45]
C Brockett, S Williams, Z M Jin, G Isaac, J Fisher. Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res B Appl Biomater 81B(2): 508-515 (2007)
[46]
V E Ozden, N Saglam, G Dikmen, I R Tozun. Oxidized zirconium on ceramic; Catastrophic coupling. Orthop Traumatol Surg Res 103(1): 137-140 (2017)
[47]
B Sagbas. Biotribology of artificial hip joints. In Advances in Tribology. P H Darji, Ed. Croatia: InTechOpen, 2016.
[48]
J B Park, C T Duong, H G Chang, A R Sharma, M S Thompson, S Park, B C Kwak, T Y Kim, S S Lee, S Park. Role of hyaluronic acid and phospholipid in the lubrication of a cobalt- chromium head for total hip arthroplasty. Biointerphases 9(3): 031007 (2014)
[49]
A M Smith, L Fleming, U Wudebwe, J Bowen, L M Grover. Development of a synovial fluid analogue with bio-relevant rheology for wear testing of orthopaedic implants. J Mech Behav Biomed Mater 32: 177-184 (2014)
[50]
A Essner, G Schmidig, A G Wang. The clinical relevance of hip joint simulator testing: In vitro and in vivo comparisons. Wear 259(7-12): 882-886 (2005)
[51]
M Parkes, C Myant, P M Cann, J S S Wong. Synovial fluid lubrication: The effect of protein interactions on adsorbed and lubricating films. Biotribology 1-2: 51-60 (2015)
[52]
D Mazzucco, G McKinley, R D Scott, M Spector. Rheology of joint fluid in total knee arthroplasty patients. J Orthop Res 20(6): 1157-1163 (2002)
[53]
A Galandáková, J Ulrichová, K Langová, A Hanáková, M Vrbka, M Hartl, J Gallo. Characteristics of synovial fluid required for optimization of lubrication fluid for biotribological experiments. J Biomed Mater Res B Appl Biomater 105(6): 1422-1431 (2017)
[54]
R Crockett, M Roba, M Naka, B Gasser, D Delfosse, V Frauchiger, N D Spencer. Friction, lubrication, and polymer transfer between UHMWPE and CoCrMo hip-implant materials: A fluorescence microscopy study. J Biomed Mater Res A 89A(4): 1011-1018 (2009)
[55]
M P Gispert, A P Serro, R Colaço, B Saramago. Friction and wear mechanisms in hip prosthesis: Comparison of joint materials behaviour in several lubricants. Wear 260(1-2): 149-158 (2006)
[56]
M Guezmil, W Bensalah, S Mezlini. Tribological behavior of UHMWPE against TiAl6V4 and CoCr28Mo alloys under dry and lubricated conditions. J Mech Behav Biomed Mater 63: 375-385 (2016)
[57]
C A Jr Engh, A S Stepniewski, S D Ginn, S E Beykirch, C J Sychterz-Terefenko, R H Jr Hopper, C A Engh. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked marathon and non-cross- linked enduron polyethylene liners. J Arthroplasty 21(6): 17-25 (2006)
[58]
Y Higuchi, T Seki, D Morita, D Komatsu, Y Takegami, N Ishiguro. Comparison of wear rate between ceramic-on- ceramic, metal on highly cross-linked polyethylene, and metal- on-metal bearings. Rev Bras Ortop 54(3): 295-302 (2019)
[59]
A Wang, A Essner, R Klein. Effect of contact stress on friction and wear of ultra-high molecular weight polyethylene in total hip replacement. Proc Inst Mech Eng H J Eng Med 215(2): 133-139 (2001)
[60]
P Damm, J Dymke, R Ackermann, A Bender, F Graichen, A Halder, A Beier, G Bergmann. Friction in total hip joint prosthesis measured in vivo during walking. PLoS One 8(11): e78373 (2013)
[61]
C Kaddick, M Malczan, C Buechele, M Hintner, M A Wimmer. On the measurement of three-dimensional taper moments due to friction and contact load in total hip replacement. In Modularity and Tapers in Total Joint Replacement Devices. A Greenwald, S Kurtz, J Lemons, W Mihalko, Eds. West Conshohocken, PA: ASTM International, 2015.
[62]
H Haider, J N Weisenburger, K L Garvin. Simultaneous measurement of friction and wear in hip simulators. Proc Inst Mech Eng H J Eng Med 230(5): 373-388 (2016)
[63]
R Sonntag, S Braun, L Al-Salehi, J Reinders, U Mueller, J P Kretzer. Three-dimensional friction measurement during hip simulation. PLoS One 12(9): e0184043 (2017)
Friction
Pages 1137-1152
Cite this article:
NEČAS D, USAMI H, NIIMI T, et al. Running-in friction of hip joint replacements can be significantly reduced: The effect of surface-textured acetabular cup. Friction, 2020, 8(6): 1137-1152. https://doi.org/10.1007/s40544-019-0351-x

885

Views

40

Downloads

28

Crossref

N/A

Web of Science

25

Scopus

0

CSCD

Altmetrics

Received: 13 August 2019
Revised: 17 October 2019
Accepted: 05 December 2019
Published: 22 February 2020
© The author(s) 2019

Open Access This article is licensed under a Creative Commons Attribution 4.0 International Li-cense, which permits use, sharing, adaptation, distribution and reproduction in any medium or for-mat, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not in-cluded in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.